6.6 Cleaning the flow-through cell

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

887 Professional UV/VIS Detector

Rinsing the flow-through cell with methanol

1 Disconnect the inlet capillary from the IC system.

2 Connect the inlet capillary to a high pressure pump or a peristaltic

pump and rinse the flow-through cell as follows – ensure that the
maximum pressure of 5 MPa is not exceeded:

■

first, rinse with ultrapure water to avoid precipitation,

■

then, rinse several minutes with methanol, to dissolve the depos-
its, and

■

finally, rinse another 15 minutes with ultrapure water, to wash
away the dissolved deposits.

3 Observe the baseline during the last rinsing step.

If the baseline remains flat, the flow-through cell is clean.

If the noise persists, rinse the flow-through cell once again with
another solvent (see "Rinsing the flow-through cell with another sol-
vent", page 32)
.

Rinsing the flow-through cell with another solvent

A solvent made up of a 1:2 mixture of acetic acid and isopropanol has
shown good results. Depending on the application, other solvents may
also prove efficient.

Preconditions:

■

Rinsing with methanol was inefficient.

■

The inlet capillary is connected to a high pressure pump or a peristaltic
pump.

1 Rinse the flow-through cell as follows – ensure that the maximum

pressure of 5MPa is not exceeded:

■

first, rinse with ultrapure water to avoid precipitation,

■

then, rinse several minutes with the selected solvent to dissolve
the deposits, and

■

finally, rinse another 15 minutes with ultrapure water to wash
away the dissolved deposits.

2 Observe the baseline during the last rinsing step.

If the baseline remains flat, the flow-through cell is clean.