Metrohm 898 XYZ Sample Changer User Manual
Page 11

■■■■■■■■■■■■■■■■■■■■■■
3 Setting up the sample changer as an RS-232 device
898 XYZ Sample Changer
■■■■■■■■
5
■
Under Program version, enter the driver version (e.g. 1.4).
■
Under Device serial number, enter the serial number of the 898
XYZ Sample Changer.
This is required in order to identify the device unambiguously. The
device serial number is to be found on the rear side of the 898
XYZ Sample Changer.
2 Set the RS-232 parameters
■
Click the tab RS 232.
■
Under COM Port, enter the connector COMX.
Select the COM port in accordance with the entry in the device
manager (see "Checking the installation", page 3).
■
Enter the following values for the RS-232 parameters.
– Baud rate 9600
– Data bit 8
– Parity None
– Stop bit 1
– Handshake None
– Terminator for send \OD (CR)
– Terminator for receive \OD (CR)
■
Click on [Connect].
The dialog window Establish connection is displayed.