MAXIM 50E/U PUMP OPERATION / MAINTENANCE MANUAL

PAGE 21

Wrap each transfer tube with 3 wraps of Teflon tape and reinstall into the

head ports using latex gloves for added grip.

Leave a protrusion height of 2”

for each transfer tube.

Position control base on mounting base, slide base forward, use locking lever

to secure.

Ensure that the ferrule of the quick grip nut is inserted into the transfer tube

port of the base. Place the gripper seal on top of the ferrule with the tapered
end facing up.

Loosely attach quick grip nuts to transfer tube ports on control base taking

care not to dislocate the ferrule and gripper seal.

Install pump onto control base by inserting transfer tubes in pump into quick

grip nuts on control base.

To secure pump to control base, tighten quick grip nuts one (1) full turn past

hand tight using 13/16” wrench.

Reinstall seals, optic probes (Probes included with E Model only), and optic

probe caps (E Model only)

into heads using the 3/4” slotted pin tool. (Do not

twist optic fibers as they could be damaged.)

Reconnect both air supply and fluid lines.