The Dynacalibrator Model 230 can be monitored and controlled remotely by
means of a serial port connection and a terminal emulation program such as
HyperTerminal.

®

The unit is controlled by entering two-character commands, shown in the table
below. Commands may be either upper case or lower case. Parameters enclosed
in brackets ([ ]) are optional. Note that for enable/disable commands, the states
indicated by an “*” are the defaults at power-up. Only the altered set point tem-
perature and function coefficients are retained when power is off and restored at
power-up.

Command

Meaning

DC[ [=] n ]

Show/set decimation count

(Data logging rate, in seconds)

L+

Start data logging

(Log data at the interval set by the decimation count:

DC=1 sets a one second interval)

L-

Stop data logging*

P+

Enable front panel control*

P-

Disable front panel control

(Prevent set point from being changed with panel controls)

SC

Show coeffecients

TE

Read environment temperature

TR

Read controlled temperature

TS[ [=] n ]

Show/set temperature setpoint

VR

Show firmware version

??

Display the list of available commands

* Factory defaults, reset at start-up

Setting Up Serial Communication via HyperTerminal

®

1. To open HyperTerminal in Windows

®

, click Start > Programs > Accessories >

Communications > HyperTerminal. The following screen will appear: