Technical Data for Mass Flow Controllers

Specification

Mass Controller

Description

Accuracy

± (0.8% of reading + 0.2% of full scale)

At calibration conditions after tare

High accuracy option

± (0.4% of reading + 0.2% of full scale)

At calibration conditions after tare

Repeatability

± 0.2%

Full scale

Operating range

1/2% to 100%

Full scale

Turndown ratio

200:1

Typical response time

100

Milliseconds (adjustable)

Standard conditions (STP)

25°C and 14.696PSIA

Mass reference conditions

Operating temperature

-10 to +50

°Celsius

Zero shift

0.02%

Full scale/°Celsius/Atm

Span shift

0.02%

Full scale/°Celsius/Atm

Humidity range

0 to 100%

Non-condensing

Controllable flow rate

102.4%

Full scale

Maximum pressure

145

PSIG

Input/Output signal digital

Mass flow, volumetric flow, pressure,

and temperature

RS-232 or RS-485 serial or
PROFIBUS or DeviceNet

Input/Output signal analog

Mass flow

0-5 VDC

Optional Input/Output signal

secondary analog

Mass flow, volumetric flow, pressure, or

temperature

0-5 VDC or 0-10 VDC or 4-20 mA

Electrical connections

8 pin mini-DIN, DB9, or DB15

Supply voltage

12 to 30 VDC

(15-30 VDC for 4-20 mA outputs)

Supply current

0.250 amp (at 12 VDC, declining with

increased supply voltage)

Mounting attitude sensitivity

None

Warm-up time

< 1

Second

Wetted materials

303 and 302 stainless steel, Viton®, silicone RTV (rubber), glass reinforced
nylon, aluminum, brass, 430FR stainless steel, silicon, glass

Advertising