VICI EUT Universal Electric Actuator User Manual

Page 14

Advertising
background image

0

Optional Serial Interface

Serial Commands

Command

Modes*

Description

AM<enter>

1, 2, 3

Displays the current actuator mode

AMn<enter>

1, 2, 3

Sets the actuator mode to [1] two position with stops, [2] two position without stops, or
[3] multiposition

CC<enter>

1, 2

Sends the actuator to Position B

CCnn<enter>

3

Sends the actuator counterclockwise to position nn (from 1 to NP)

CNT<enter>

1, 2, 3

Displays the current value in the actuation counter

CNTnnnnn<enter>

1, 2, 3

Sets the actuation counter from 0 to 65535.
(For example, to reset the counter, type CNT0<enter>)

CP<enter>

1, 2, 3

Displays the current position

CW<enter>

1, 2

Sends the actuator to Position A

CWnn<enter>

3

Sends the actuator clockwise to position nn (from 1 to NP)

DELn<enter>

1, 2, 3

Sets the command delimiter to [1] CR (default), [2] LF, or [3] CRLF

DT<enter>

1, 2

Displays the current delay time in milliseconds

DTnnnnn<enter>

1, 2

Sets the delay time from 0 to 65,000 milliseconds

GOnn<enter>

1, 2

Sends the actuator to position n, where n is A or B

3

Sends the actuator to position nn (from 1 to NP) via the shortest route

HM<enter>

3

Moves the valve to postion 1 (home)

ID/<enter>

1, 2, 3

Sets the ID of the actuator to / (Must be 0-9 or A-Z)

*ID*<enter>

1, 2, 3

Resets the ID to None

IFMn<enter>

1, 2, 3

Sets the interface response mode to [0] no response string when an action command
is sent, [1] basic response to action commands, or [2] extended response to action
commands (needed for BCD interface)

LGn<enter>

1, 2, 3

Sets the interface response mode to [0] no extra characters, or [1] longer response string to
be fully backwards compatible with VICI microelectric actuators

LRN<enter>

1

Forces the actuator to find the stops on a newly installed valve.
(Note: The valve must be installed before this command is sent.)

NP<enter>

1, 2, 3

Displays the number of positions the actuator is currently set to index

NPnn<enter>

2

Sets the number of ports (nn) for the current valve. (Must be an even number between
2 and 40)

3

Sets the number of positions (nn) for the current valve. (Must be an even number between
2 and 40)

SB<enter>

1, 2, 3

Displays the current baud rate

SBnnnn<enter>

1, 2, 3

Sets the baud rate to 48(00), 96(00), 192(00), 384(00), 576(00), or 1152(00). The parity setting,
number of data bits, and number of stop bits cannot be changed.

SM<enter>

3

Displays the current default rotational direction

SMl<enter>

3

Sets the default rotational direction to [F] for forward rotation, [R] for reverse rotation, or
[A] to automatically choose the shortest route.

SO<enter>

1, 2, 3

Displays the current offset value

SOnn<enter>

1, 2, 3

Sets the offset value of the first position to be any number from 1 to 96 minus the total
number of positions. Example: for a 10 position valve, the offset can be set from 1 to 86.

STAT<enter>

1, 2, 3

Displays the status of the actuator

TM <enter>

1, 2, 3

Displays the amount of time required for the previous move, in milliseconds

TO<enter>

1, 2

Toggles the actuator to the opposite position

TT<enter>

1, 2

Toggles the actuator to the opposite position, waits a preset delay time, then rotates back
to the original position.

VRn<enter>

1, 2, 3

Displays the current firmware version for [NULL], the main PCB, or [2] the serial interface PCB.

/?<enter>

1, 2, 3

Displays a list of valid commands

Modes are described on pages 4-5.
Notes:

In this chart, nn represents numbers to be entered and l represents letters.

<enter> = CR(0x0D)

*