With serial control – VICI EUT Universal Electric Actuator User Manual

Page 24

Advertising
background image

0

With Serial Control

Here is how this can be feature can be used to set up a 31-stream stream selection system
using two 16-position valves and two actuators serially daisy-chained together. Refer to
the chapter entitled “Optional Serial Interface” (beginning on page 6) as required.

1. Configure the second actuator using the command SO16, giving it a valid position

range of 16 to 31.

2. Use a piece of tubing to connect port 16 of the first valve (on the actuator still carrying

the factory-default offset valve of “1”) to the common port of the valve on the second
actuator (which now has an offset value of “16”).

3. Connect streams 1 through 15 to ports 1 through 15 on the first valve, and streams 16

through 31 to ports 1 through 15 on the second valve.

Using the Offset Feature

Advertising