Weller WCB1 User Manual

Page 2

Advertising
background image

2

TEMP
Function key is not used.
OPERATING INSRUCTIONS

ERRORS
ERR:

General error (communication error)

BAT:

Indicates that battery must be replaced

HI:

Entry is too high

LO:

Entry is too low

F01:

Error in calibration (lower value)

F02:

Error in calibration (upper value)

No blinking cursor in ready mode:

Error in connection cable (5) (defective or not connected)

SOLDERING TOOL CONNECTION
For operating with soldering tool, the soldering tool can be inserted in socket (2).
AUTOMATIC SWITCHING OFF
If the calibration box is not connected with the power unit, it will be switched off after 2 minutes.
BATTERY REPLACEMENT
Push open cover to battery compartment in direction of arrow (open). Insert batteries as shown in battery compartment. Shut housing.
TECHNICAL DATA
Battery 4 x 1.5V Type AA
SCOPE OF SUPPLY
5 31 182 99 Calibration box
Batteries 4 x 1.5V
Operating Instructions (only valid for WCB1 USA version!)