Great Plains YP1625A-1670 16 Row 70 cm Quick Start User Manual

Page 3

Advertising
background image

Quick Setup Guide for IntelliAg Model YP1625 70 cm Air Pro

11001-1603-201108

©2011 DICKEY-john Corporation
Specifi cations subject to change without notice.

TABLE E:

Planter Control

Setup

Default Value/

Value to Enter

Split Air Regulation

Instructions/Defi nitions

Type

Split Air Reg

Confi gure Control Channel 2 as Split Air Regulation.

Control Mode

Auto

Control channel feedback based on air pressure sensor.

Drive Type

Air Reg 2

Automatically selects Air Reg 2 as drive type.

Input Filter

79

Amount of fi ltering applied to the control channel feedback frequency.

Pressure Drop

0

Difference in pressure sensor mounting location to the seed disk in inches

of H20.

Pressure Slope

447.2269

The change in pressure sensor voltage to a frequency readable by the

IntelliAg system and measured in inches of H20. Should only be adjusted

by qualifi ed personnel.

Pressure Offset

800

Takes a zero point reading that provides a frequency when the fan is off.

Press the Zero Pressure button to calculate frequency.

Planter Selection

YP16

Select planter selection of YP1625 automatically adjusts to the appropri-

ate calibration settings.

Sensitivity Adjust

0

Increases or decreases the calibration parameters in the ranges of -10 to

+10. Increasing response time makes the system response quicker.

STEP 7B: Planter Control Channel Setup
(Split Air Regulation)

NOTE: Split Air Regulation must be confi gured as Control Channel 2
only.

At the Control Setup screen, press the Channel Setup button

.

Select Channel 2 and select Split Air Regulation as the Type.

Enter desired values from Table E.

Press the Work Screen button

to return to the Main Work screen.

1.

2.

3.

4.

STEP 7A: Material Confi guration Setup (Split Air Regulation)

It is recommended that when setting a control channel for split air, the
material name be created as “Air” to eliminate confusion between the
actual material and the control used.

At the Main Menu screen, press the Control Setup button

.

Press Material button 16.

Enter desired values from Table D.

Press the Work Screen button

to return to the Main Work screen.

1.
2.

3.

4.

TABLE D:

Material Setup

Default Value/

Value to Enter

Split Air Regulation

Instructions/Defi nitions

Type

Split Air Reg

Desired type of application control channel being used for a specifi c mate-

rial. CREATE MATERIAL NAME AS “AIR”.

Units

In H20

Oz in2

Automatically changes with the type of material application selected.

Changes units for target application.

Preset Method

Disabled

User-defi ned target rates can be confi gured and when enabled can be ad-

justed from the Main Work screen using the Increment/Decrement buttons.

Target Rate

2.00

Establishes the desired rate of application in inches of H20.

Max Rate

5.00

Maximum application rate in inches of H20.

Min Rate

1.00

Minimum application rate in inches of H20.

Inc/Dec %

5%

Percentage of change of the entered target rate applied each time the

Increment/Decrement button is pressed on the Main Work screen.

STEP 8: Row Monitor Setup

At the Main Work screen, press the Row Monitor button

.

Enter desired values using Table F as reference.
Press the Work Screen button

to return to the Main Work screen.

1.
2.
3.

TABLE F:

Row Monitor

Setup

Default Value/

Value to Enter Instructions/Defi nitions

Material Name

See Instructions

Material Name only appears on the Row Monitor Setup screen when all

control channels are disabled and material is set for Monitor Only. This is

only used for ground drive/nonhydraulic applications to monitor population

with high and low alarms. A material must be confi gured and selected to

activate alarms.

High Alarm Delay

5

Desired number of seconds that high population can be above high alarm

point before alarm will sound.

Low Alarm Delay

5

Desired number of seconds that low population can be below low alarm

point before alarm will sound.

Population Adjust

100

Enter a % to allow for seed sensor population inaccuracies to achieve the

desired population display. 100% represents true calculation.

Population Filter

50

Set fi lter value to stabilize the monitored population display. Number can

be set to 0% for no fi ltering and 99% for high level fi ltering.

Row Fail Rate

2/1 (2 seeds

every 1 second)

Set to desired number of seeds per second to trigger seed sensor failure

alarm.

3

Advertising