Parameter Settings

6-29

Height h

Mounting height water-US int + pressure int

Height H

Mounting height water-US NIVUS

Mounting offset of level (depth) sensors

After modifying the mounting height of the pressure or water-
ultrasonic depth sensors, you must adjust the mounting height in the
PAR > Flow velocity menu by the same value!

Select Layers

This parameter will display only when you have selected a sensor combination. The
Portable FlowShark Pulse automatically aligns the sensors to partial layers. However,
you can define the borders for these layers manually using the ALT key. Designate
the threshold levels between the layers using the boxes in the from lines.

1

Sensor corresponding to the top layer

2

Threshold level between middle and top layers

3

Sensor corresponding to the middle layer

4

Threshold level between middle and bottom layer

5

Sensor corresponding to the bottom layer

Selecting the layers

After selecting the sensors, they will display on the screen.