ADS Environmental Services Portable FlowShark Pulse QR 775006 A2 User Manual
Page 54

Installation 3-17
Proper installation (left) and improper installation (right)
Installation susceptible to buildup of debris (left) and improper installation (right)
Sensor properly facing flow direction (left) and sensor facing the wrong direction (right)
Please note the following for installation in pipes:
• Weld on the 1.5-inch nozzle at a 90° angle.
• Position the sensor so that the bevelled edge is facing
upstream.
The sensor block must be welded to steel and stainless steel 1.4571, glued to PVC,
welded to HDPE, or laminated to PVC. To perform upgrades, ADS recommends
using a tapping saddle. Make sure the access hole into the pipe is 1.5 inches (38 mm)
in diameter and goes completely through the pipe (including any ball or gate valve) to