5-8

ADS Portable FlowShark Pulse Manual

t1

=

Programming time (any period)

t2

=

System reset and restart (approximately 7 seconds)

t3

=

Cycle time (constant; will change only for designated events; 1 minute…60
minutes)

t4

=

Measurement duration, depending on hydraulic and physical conditions,
will reset each time (5 seconds…40 seconds)

Measurement and display functions following parameter modification

Display Functions Outside Memory Mode

When initially setting up the Portable FlowShark Pulse system for operation in
difficult applications, the memory function may not be necessary for short-term
projects and scheduled verification of other monitoring systems (e.g., flumes, weirs,
or magnetic-inductive systems) or throttles. However, it may be beneficial to display
the current readings real-time. Therefore, the unit will continue to take and display
measurements even when the memory function is disabled.

The current readings will display continually on the unit, but these
readings will be saved only if the memory mode has been enabled.
In addition, please note that exercising this option significantly
increases power consumption.