Filter replacement – Bio-Rad S3™ Cell Sorter User Manual

Page 22

Advertising
background image

12 | S3 Biosafety System

Maintenance

Filter Replacement

Replacement of the HEPA filter requires proper training and should be performed by the
certifying agent. Once the HEPA filter is changed the S3 biosafety system must be re-certified
to ensure proper safety.

Please contact a certified biosafety service vendor when the filter life reaches 10% to schedule
replacement.

1. Completely shut down the S3 or S3e system and power off.
2. Power off the biosafety system main power unit and unplug.
3. Unplug the USB cable.
4. Unclamp the upper filter housing.
5. Tilt the front upward. This should expose the luer fitting on the sensor tubing. Separate

the fitting.

6. Completely remove the upper filter housing and set aside.
7. Lift the exposed HEPA filter from the housing and dispose of it, following facility protocol.
8. Check and clean the inside of the filter chamber and sterilize as needed.
9. Slide the new HEPA filter into the housing with the flow arrows pointing up.
10. Replace the upper filter housing with the sensor hose in the front.
11. Tilt up the upper housing, exposing the hose, and reattach the luer fitting.
12. Compress the filter gaskets with the four draw latches.
13. Attach the USB cable.
14. Attach the power cable and switch on.
15. Power on the S3 or S3e system and check the filter status and air speed for the

specified values.