Service and certification, Service onsite certification – Bio-Rad S3™ Cell Sorter User Manual

Page 23

Advertising
background image

Instruction

Manual | 13

Whenever service is conducted on the biosafety system, be sure to completely power down
both the S3

or S3e cell sorter through the main power switch and the biosafety system.

Service

Bio-Rad offers service support for the biosafety system. Contact Bio-Rad’s technical and service
support toll free at 1-800-4BIORAD (or 1-800-424-6723) for service needs or replacement parts.

Bio-Rad does not conduct certification of the system onsite (this must be done by third party
service organization), but upon installation a system test will check for proper airflow. The S3
biosafety system will also be tested to ensure connectivity with ProSort

software.

Onsite Certification

As with any biosafety enclosure or hood, the S3 biosafety system must be certified upon
installation and annually to maintain proper safety. The system cannot be used until it is
certified onsite.

Certification should be performed by a qualified technician in accordance with the attached
procedure. Contact Bio-Rad Technical Support for recommended vendors that can
certify onsite.

CAUTION: Risk of danger! This procedure does not address the possible exposure
to potentially hazardous substances. Check with a biosafety professional or a health
and safety professional before performing any service on this enclosure, including
repair or replacement of the HEPA filter.

Service and Certification

4