S3 biosafety system class i certification form – Bio-Rad S3™ Cell Sorter User Manual

Page 26

Advertising
background image

Service and Certification

16 | S3 Biosafety System

S3 Biosafety System Class I Certification Form

S3 Biosafety System Class I Certification Form

Customer Name

Phone

Company

Location

Enclosure SN

Inflow Velocity

Airflow readings, FPM __________________________

__________________________

__________________________

__________________________

__________________________

__________________________

__________________________

__________________________

__________________________

__________________________

Calculated average airflow

_________________________ FPM

Average airflow >135 FPM?

______ PASS or ______ FAIL

Measured area

______ ft

2

Area of perimeter base slots

______ 0.15 _________ ft

2

Total area (measured area + perimeter slots)

______ ft

2

Exhaust volume (average airflow x total area)

_________________________ CFM

Leak Integrity

Upstream concentration

______________________________

Upstream concentration >10 µg/L?

______ PASS or ______ FAIL

Exhaust discharge concentration

______________________________

Exhaust discharge <0.01% upstream concentration?

______ PASS or ______ FAIL

Smoke Containment

Smoke drawn into enclosure?

______ PASS or ______ FAIL

Certifier (Printed Name)

Certifier (Signature)

Date of Certification

Advertising
This manual is related to the following products: