Bio-Rad Clarity™ Western ECL Substrate User Manual

Page 11

Advertising
background image

Technical Support: 1-800-424-6723 • www.bio-rad.com 7

4. Remove the membrane from the substrate solution and

drain off excess.

5. Place membrane in a plastic sheet protector or in plastic

wrap to prevent the membrane from drying.

6. Image the membrane with a digital imager or by exposing to

X-ray film.

If switching to Bio-Rad Clarity from another chemi
substrate, optimal exposure times may be different.
Refer to the Substrate Transition Guide below when
switching substrates

Transition Guide. To transition to Clarity, use the following
guidelines to adjust exposure times.

Current Substrate

Exposure Using Clarity

GE ECL/Pierce ECL

Decrease 20-fold*

SuperSignal West Pico

Decrease 5-fold*

Bio-Rad

®

Immun-Star

HRP

Decrease 5-fold*

Pierce ECL Plus

Equal

Bio-Rad

®

Immun-Star

WesternC

Equal

SuperSignal West Dura

Equal

ECL Prime

Increase 2-fold

SuperSignal West Femto

Increase 2-fold

ECL Select

Increase 2-fold

*

Or decrease antibody dilution by 2- to 5-fold.

ECL Select is a trademark of GE Healthcare UK limited.