8 Technical Support: 1-800-424-6723 • www.bio-rad.com

Troubleshooting

Problem

Cause

Solution

High

background

Blocking was

incomplete

Increase the

concentration of blocking

agent or increase

blocking duration

Match the blocker to

the membrane. For

example, gelatin may give

poor results on PVDF

membranes

Washing was

insufficient

Increase the number,

duration, or stringency of

the washes

The primary

or secondary

antibody

was too

concentrated

Decrease antibody

concentrations

Perform a dot-blot

experiment to

optimize the working

concentrations

Areas of no

signal within

a band

(a donut

appearance)

Localized

substrate

depletion

Bands on the blot with

high protein amounts

will lead to excessive

local concentrations of

peroxidase conjugate.

This may lead to localized

substrate depletion. Use

lower concentrations of

primary and secondary

antibody

Load less sample