Bio-Rad Quantity One 1-D Analysis Software User Manual
Page 301
Quantity One User Guide
D-12
Fig. D-10. Step Tablet dialog box.
The dialog box will indicate whether the displayed form is for the transmissive or 
reflective step tablet, depending on whether you are in transmissive or reflective 
mode.
Note:
In transmissive scanning, the scanner is calibrated against a transmissive step 
tablet. In reflective scanning, the scanner is calibrated against a reflective step 
tablet, unless the reflective calibration strip is uncalibrated. If the reflective 
calibration strip does not have step tablet values associated with it, you can use 
the default values in the form.
In the Tablet Serial Number field, you can type in the serial number for the step tablet 
values you are entering. (The reflective step tablet will probably not have a serial 
number.)
Your calibration strip will have a clear plastic cover. If the quantity offset value for 
the clear cover is included in your calibration strip package, you can enter this 
number in the Quantity Offset field. Otherwise, use a value of 0.045.