Clear bubbles routine, E (see, Clear bubbles routine on – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual
Page 79
Advertising

Clear Bubbles Routine
User Guide
| 77
Clear Bubbles Routine
Description of Routine
The Clear Bubbles routine performs a series of fluidics operations to force out air
bubbles that may be trapped in the cuvette.
When to Run this Routine
If you suspect there is air in the MAGPIX system, run the Clear Bubbles routine.
Estimated Run Time
3 minutes
To Run the Clear Bubbles Routine
1.
Navigate to the Maintenance view.
2.
In the Select Routine section, select Clear Bubbles from the Additional
Routines dropdown list.
3.
Add 5 ml distilled or deionized water and 5 ml 70% isopropanol to the
appropriate reservoirs in the reagent block.
4.
Click Start.
Advertising
This manual is related to the following products: