Unclog Routine
Description of Routine
This procedure performs a series of fluidics operations to remove possible
obstructions from the fluidics lines.
When to Run this Routine
If you suspect there is an obstruction in the MAGPIX system, run the Unclog routine.
Estimated Run Time
2 minutes
To Run the Unclog Routine
1.
Navigate to the Maintenance view.
2.
In the Select Routine section, select Unclog from the Additional Routines
dropdown list.
3.
Add 5 ml distilled or deionized water and 5 ml 0.1N NaOH to the appropriate
reservoirs in the reagent block.
4.
Insert the plate in the microplate platform and click Start.