Shut Down Routine
User Guide
| 83
Shut Down Routine
Description of Routine
The Shut Down routine cleans the fluidics lines and prevents a buildup of debris
within the system.
When to Run the Routine
Bio-Rad recommends that you run the Shut Down routine at the end of each day
that you use the instrument. Refer to the Shut Down entry in the status section in
the Maintenance view to determine whether the Shut Down has been run that day.
Estimated Run Time
2 minutes
To Run the Shut Down Routine
1.
Navigate to the Maintenance view.
2.
In the Select Routine section, select Shut Down.
3.
Add 5 ml distilled or deionized water and 5 ml bleach to the appropriate
reservoirs in the reagent block.
4.
Click Start.