Microplate requirements, Recommended microplates for magnetic assays, Recommended microplates for vacuum filtration – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 12

Advertising
background image

Basic Operation and Maintenance

12 Bio-Plex Pro and Bio-Plex Pro II Wash Stations

Microplate Requirements

Use recommended microplates. If you use different microplates, your
assay results may not be reproducible, and you may experience
excessive bead loss.

Recommended microplates for magnetic assays:

Bio-Rad Part #

Description

Default for Magnetic Products

171-025001

Bio-Plex Pro

flat-bottom plates

Yes

*For a complete list of suitable plates, please contact Bio-Rad technical support.

Only these recommended filter plates should be used with the Bio-Plex Pro II, when configured for standard
Bio-Plex or other polystyrene bead based xMAP assays.

Recommended microplates for vacuum filtration:

Vendor

Manufacturer Part #

Default for Vacuum

Products

Millipore MSBVS1210

Yes

If using other microplates, you will have to enter new microplate definitions, using the
Hydrocontrol software included on the CD. Please note the following requirements.

Parameters Characteristics

Max. overall microplate height

14.35 mm ± 0.76 mm
(0.5650 inches ± 0.0299 inches)

Footprint

9.0 mm (0.3543 inches)

Pitch size (center to center)

9.0 mm (0.3543 inches)

Bottom shape

Round, V-shaped, and flat

Advertising