Maintenance and cleaning, Cleaning procedures, Cleaning the cover and display – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 65: Cleaning the liquid system

Advertising
background image

Troubleshooting and Error Messages

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

65

Maintenance and Cleaning

Cleaning Procedures

WARNING

All parts of the instrument that come into contact with potentially
infectious material must be treated as potentially infectious areas.

It is advisable to adhere to applicable safety precautions, (including the
wearing of powder-free gloves, safety glasses and protective clothing) to
avoid potential infectious disease contamination when performing
cleaning procedures and also when making adjustments to the
instrument.

The most important cleaning procedure for this instrument is to rinse the liquid system with distilled
water before the instrument is left to stand, or switched off at the end of each day.

The manifold should be removed and cleaned thoroughly at least once every six months, or
whenever one or more of the needles are blocked.

Cleaning the Cover and Display

The outer surface of the instrument and the display may be cleaned periodically
using a tissue moistened with a mild detergent solution.

Caution

Never use acetone, as it will damage the covers.

Cleaning the Liquid System

To clean the liquid system, perform Rinse and Prime procedures as described in the
Basic Operation and Maintenance section.

Caution

If the manifold is not rinsed, the needles will become blocked.
If this occurs, the manifold will need expensive repairs or will have to be
replaced.

Advertising