Defining New Programs

56 Bio-Plex Pro and Bio-Plex Pro II Wash Stations

Aspiration Position Diagrams

Position Bottom

Position Custom

Position Overflow

Aspiration Position Diagrams

Process Step: Dispense

Select required dispensing position
(OVERFLOW, CUSTOM, MOVE OVER.,
MOVE CUST. or BOTTOM).

Select the channel to be used (1 - 4)
depending on instrument configuration.

Select the volume of solution to be used
(50 - 400 µl in increments of 50 µl).

Select and confirm dispensing step (DISP)
CX = Cycle Number
PX = Program Step Number

CX PX DISP

- + EXIT OK

VOLUME 300µl

- + EXIT OK

CHANNEL 1

- + EXIT OK

FL.RATE 300 µl/s

- + EXIT OK

Z-POS: OVERFLOW

- + EXIT OK

Select required flow rate (drip to 500µl/sec).

After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another dispense step
or press – or + to select the next program step,
Program End or a new cycle.

CX PY DISP

- + EXIT OK