Bio-Rad Human MMP and TIMP Assays User Manual

Page 29

Advertising
background image

27

Luminex xPONENT Software
Luminex xPoNENT software may be used to analyze Bio-Plex assays.
Although guidelines are provided here, consult the xPoNENT software
manual for more details. Perform a system initialization with Luminex’s
calibration and performance verification kit, as directed by Luminex. Select
Batches to set up the protocol and follow the information under Settings.

Note: The instrument settings described below apply to Luminex 100/200
and FLEXMAP or Bio-Plex 3D instruments. For the Bio-Plex MAGPIX
reader, use the default instrument settings.

1. Select MagPlex as the bead type for magnetic beads. This

automatically sets the DD gates.

2. Volume = 50 µl.

3. Refer to Table 11 to select the appropriate PMT setting for your

instrument.

4. Plate name: 96-well plate.

5. Analysis type: Quantitative, 5PL Curve Fit.

6. Number of standards: 8.

Select Analytes to set up the panel.

1. Enter “ng/ml” in the Units field.

2. Enter 50 in the Count field.

3. Select the bead region and enter the analyte name.

4. Click Apply all for Units and Count.

Select Stds and Ctrls.

1. Enter standard concentrations, lot number, dilution factor, and other

information as applicable.

After the assay is complete, select Results, then select Saved Batches.