Prepare coupled beads, Prepare coupled beads 16 – Bio-Rad Bio-Plex Pro™ Human Cancer Biomarker Assays User Manual

Page 18

Advertising
background image

16

6. Prepare Coupled Beads

1. Use Tables 6–7 or the Calculation Worksheet on page 35 to calculate

the volume of coupled beads and assay buffer needed.

2. Add the required volume of Bio-Plex

®

assay buffer to a 15 ml

polypropylene

tube.

3. Vortex the 20x stock of coupled beads at mid speed for 30 sec.

Carefully open the cap and pipet any liquid trapped in the cap back
into the tube. This is important to ensure maximum bead recovery.
Do not centrifuge the vial; doing so will cause the beads to pellet.

4. Dilute coupled beads to 1x by pipetting the required volume into the

15 ml tube. Vortex.

Each well of the assay requires 2.5 μl of the 20x stock adjusted to a

final volume of 50 μl in assay buffer.

5. Protect the beads from light with aluminum foil. Equilibrate to room
temperature prior to use.

Note!

To minimize volume loss, use a 200–300 μl capacity pipet to remove

beads from the 20x stock tube. If necessary, perform the volume transfer
in two steps. Do not use a 1000 μl capacity pipet and/or wide bore pipet tip.

Preparing 1x coupled beads from 20x stock (includes 20% excess volume)

Table 6. Premixed panel or one singleplex assay.

Table 7. Mixing singleplex assays.

# of Wells

20x Beads, µl

Assay Buffer, µl

Total Volume, µl

96

288

5,472

5,760

48

144

2,736

2,880

20x Beads, µl

20x Beads, µl

# of Wells

Singleplex #1

Singleplex #2

Assay Buffer, µl

Total Volume, µl

96

288 288 5,184 5,760

48

144 144 2,592 2,880

Advertising