7. Run Assay

The following instructions apply to premixed multiplex, singleplex, x-Plex

,

and Express assay formats.

Considerations

n

Bring all assay components and samples to room temperature before use

n

Use calibrated pipets and pipet carefully, avoiding bubbles

n

Pay close attention to vortexing, shaking, and incubation instructions.

Deviation from the protocol may result in low assay signal and
assay variability

n

Assay incubations are carried out in the dark. Cover the plate with

aluminum foil or otherwise protect from extended exposure to light

Table 8. Summary of wash steps and incubations. After each assay step, select the

appropriate Bio-Plex Pro

wash station program or perform the appropriate manual wash step

as summarized below.

Using a Vacuum Manifold

n

After each incubation, place the filter plate on a calibrated vacuum

apparatus and remove the liquid by vacuum filtration

n

To wash, add 100 μl wash buffer to each well and remove the liquid as

before. Ensure that all wells are exposed to the vacuum

n

Thoroughly blot the bottom of the filter plate with a clean paper towel

between each vacuum step to prevent cross contamination

17

Bio-Plex Pro or

Bio-Plex Pro II

Handheld Magnet or

Pro II Wash Station

Wash Station

Vacuum Manifold

Assay Step

Magnetic Program

Vacuum Program

Manual Wash Steps

Add beads to plate

MAG x2

VAC x2

2 x 100 μl

Sample incubation
Detection Ab incubation

MAG x3

VAC x3

3 x 100 μl

SA-PE incubation