1. General Information

1.1

Introduction

The Benchmark Plus microplate reader provides all the flexibility required for absorbance assay.

The Benchmark Plus has the capability to perform standard colormetric ELISA and kinetic

readings for general cell screening with 6, 12, 24, 96, and 384-well microplate formats. The

Benchmark Plus uses a “filter-less” design, with a wavelength range of 340–800 nm. Benchmark

Plus is also provided with an optional incubation feature with temperature range up to 45

degrees Celsius. The optical design provides verifiable results up to 4.0 O.D. The “filter-less”

monochromator design allows the user to set the reading at any wavelength in the 340 – 800 nm

region, as well as do a spectral scan across any wavelength range (within 340 – 800 nm) to find

the absorbance curve.

1.2

Benchmark Plus Specifications

Light Source

Xenon Flash Lamp

Spectral Range

340–800 nm

Indication range

0 – 4.000 OD

Resolution

0.001 OD

Monochromator Capacity

Any wavelength within the spectral range, with 1 nm step size

Monochromator Bandwidth

5 nm

Wavelength Accuracy, min.

±

2.0 nm

Wavelength repeatability

±

0.2 nm

Shaking

Yes

Optional Incubation

Up to 45

°

C , programmable in 0.1

°

C increments

Power

100–240 V, 60/50 Hz, 2.0 Amps

Dimensions

254 mm (H) x 457 mm (W) x 508 mm (D)

Compatible Plates

Any 6- through 384-well microplate with dimensions 138 mm (length) x 98
mm (width). Plate height cannot exceed 15 mm.

Linearity (340 – 800 nm)

Less than 2% or 0.01 OD from 0–3.0 O.D.

Accuracy (340 – 800 nm)

Less than 1% or 0.01 OD from 0–3.0 O.D.

Reproducibility (340 – 800 nm)

Less than 1% or 0.005 OD from 0-3.0 OD

Software Compatibility

A System Pentium 166 or higher, 32 Mb RAM, Windows 95 OS or higher

Minimum Kinetic Read Interval

96-well plate:

10 seconds single wavelength

30 seconds dual wavelength

Environmental Operating
Range

5–45 degrees C, max. relative humidity 80% non-condensing