8

Microplate Manager (see the Microplate Manager User Guide, section 2.7). Incorrect
wavelength and clip information could result in damage to the Ultramark!

Make sure that all the filters sit securely and properly in the Magazine clips before reinstalling the
Magazine.

Reinsert the Filter Magazine into the base so that the pins in the base fit in the holes in the magazine.
Insert Screw 2 into the center hole in the Filter Magazine and tighten securely with a coin or
screwdriver.

The Filter Magazine must be securely fastened to the base to avoid serious problems.

Replace the side cover and fasten it with the thumbscrews. Restart the Ultramark.

Finally, be sure to enter the new filter information into Microplate Manager (see section 2.7 of the
Microplate Manager User Guide).

Incorrect

Correct