Bio-Rad Model 550, Ultramark, and Benchmark Microplate Reader Accessories User Manual

Page 23

Advertising
background image

17

Error Message

Probable Cause

Recommended Solution

SCSI ini.

SCSI chip trouble.

Contact Bio-Rad.

WCS R/W SCSI

SCSI chip trouble.

Contact Bio-Rad.

DPR R/W SCSI

SCSI chip damaged.

Contact Bio-Rad.

SCSI bus

Data communication problem
between the Ultramark and
your computer.

The Ultramark is not the source of the
problem. Check your computer.

Bus Rst SCSI

The termination switch may be
off.

Turn the SCSI termination switch on
(located next to SCSI connector).

Another connected SCSI
device is powered off.

Turn the connected machine on, or
turn the termination switch on the
Ultramark on.