Bio-Rad Aminex Organic Acid and Alcohol Analysis Columns User Manual

Page 18

Advertising
background image

16

6.5 Rectifying Instrument/Connecting Tubing Problems

Many instances of poor resolution are not the result of column problems, but rather
the result of hardware problems elsewhere in the system. The following list can
help to point out troublesome areas and provide ideal specifications.

1.

Injector

Sample volume

100 µl maximum

Loop volume

500 µl maximum

2.

Connecting Tubing

Fittings not deteriorated
Tube end cut square
Tube bottoms in fitting
Tube ID

0.013” maximum

Tube length

10” maximum

3.

Detector

Clean cells
Cell inlet tubing 0.013” ID maximum
Column connected to sample side, not reference side
Temperature stable

Section 7
Heated Column Shut-Down

After a high temperature separation, reduce the flow rate to 0.2 ml/min and turn off
the column heater. Continue to pump solvent until the column returns to ambient
temperature. Removing a hot column from the system will cause the solvent to
contract while cooling and air will be pulled into the column.

If a fast change between two columns is required, the hot column can be
immersed in filtered, fresh distilled deionized water to cool rapidly. Remember to
replace the end screws tightly to prevent the resin from drying out.

Lit42D.qxd 11/29/2005 2:23 PM Page 16

Advertising
This manual is related to the following products: