Table 3. Troubleshooting Guide

Problem

Cause

Characteristics

High back

1. Bed collapse from excessive Sudden catastrophic pressure

pressure

flow rate

increase at flow above
recommended maximum

2. Chemical contamination

Gradual increase in pressure
during use

3. Trapped air

Column returned to use
after storage

4. Microbial contamination

Column returned to use
after storage

5. Change in ionic form

Rapid pressure increase after
buffer change

6. Bed degradation from

Sudden catastrophic pressure

wrong organic solvent

increase with solvent change

Loss of resolution

1. Particulate contamination

Gradual loss in efficiency or
changing retention

2. Change in ionic form

Changing retention times;
Sudden change which may be
accompanied by poor efficiency

3. Bed collapse from excessive Retention times constant; poor

flow rate

efficiency; skewed peaks.
Sudden change.

4. Bed compression

Retention times constant;
lowered efficiency; tailing
skewed peaks. Gradual change.

5. Bed degradation from wrong Changing retention times

`

organic solvent

accompanied by poor efficiency

6. Excessive dead volume

Inability to match test
chromatogram

Lit42D.qxd 11/29/2005 2:23 PM Page 18