Bio-Rad Profinity Epoxide Resin User Manual

Page 8

Advertising
background image

6

Recommended Columns
Bio-Rad’s Bio-Scale

MT high-resolution columns may be used for the above

column packing procedure. These columns are empty but may be packed with
the support of choice. Bio-Scale MT columns are convenient for use with
Bio-Rad’s BioLogic

system or with any medium- or high-pressure system.

Bio-Scale MT2 column (7 x 52 mm) for bed volumes up to 2 ml

Bio-Scale MT5 column (10 x 64 mm) for bed volumes up to 5 ml

Bio-Scale MT10 column (12 x 88 mm) for bed volumes up to 10 ml

Bio-Scale MT20 column (15 x 113 mm) for bed volumes up to 20 ml

Column packing — sample preparation sized columns

Use this method for packing Profinity resin into small micro spin columns for
sample preparation (such as Bio-Rad’s Micro Bio-Spin

chromatography

columns, empty, 100, catalog #732-6204).

1.

Make a 50% slurry of ligand-coupled Profinity epoxide resin in degassed
binding buffer of choice. Resuspend the slurry by gently swirling or stirring
with a glass or plastic rod. Do not use a magnetic stirrer.

2.

Place the column into an appropriate collection vessel (for example, a 2 ml
capless collection tube), and spin.

3.

Using a pipet, transfer enough Profinity resin to a microcentrifuge tube. If
using Bio-Rad’s Micro Bio-Spin column, transfer ~0.2 ml slurried Profinity
resin to a Micro Bio-Spin column. This is equivalent to ~100 µl of a packed
resin bed.

4.

Remove excess binding buffer by centrifugation. Centrifuge at 1,000 x g for
15 sec to pack resin. The column is now ready for separation.

Recommended columns for gravity-flow chromatography

The following Bio-Rad columns may be used for gravity-flow chromatography.
They are empty and can be filled with Profinity resin.

Poly-Prep

®

columns, for up to 2 ml chromatography support and 10 ml

sample

Econo-Pac

®

columns, for up to 20 ml

Glass Econo-Column

®

columns, from 5–170 cm long and 0.5–5.0 cm in

diameter