Section 9
Troubleshooting Guide

Problem

Possible Cause

Possible Solution

Insufficient negative Open vacuum regulator

Close vacuum regulator

pressure in vacuum

by turning completely

manifold

clockwise

Manifold top not seated

Ensure uniform contact

properly on base

between manifold top
and base

Purification plate or CAP

Press down gently on

not fully seated

the corners of the plate
or CAP to seal

Unused CAP luer fittings

Close unused luer

not closed completely

fittings completely with
caps

Gasket residue preventing Rinse manifold with
even contact between

distilled water

manifold parts

Lower gasket misaligned

Press gasket back into
groove

Physical damage to

Replace vacuum

manifold or CAP (e.g.,

manifold or CAP

chips and cracks)

Purge step in purification

If closure of wells or

protocol

columns (e.g., with
sealing tape) restores
negative pressure, no
corrective action
needed

Excessive noise

Purification plate or CAP

Press down gently on

during vacuuming

not fully seated

plate or CAP to seat

Gasket residue preventing Rinse manifold with
even contact between

distilled water

manifold parts

Worn manifold gasket

Replace gasket

Problem

Possible Cause

Possible Solution

Incorrect negative

Vacuum source

Connect vacuum

pressure from

connected to left

to right side of vacuum

vacuum regulator

side of vacuum

Clogging of

Insufficient negative

See “Insufficient

purification plate

pressure in vacuum

negative pressure

or CAP columns

manifold

in vacuum manifold”
problem above