Problem

Possible Cause

Recommended Solution

Low yield

Insufficient amount of

Scale up the amount of

(continued)

PureZOL used for

PureZOL used according to

sample lysis and

sample size. Sample volume

homogenization

should not exceed 10% of
the PureZOL reagent used
for homogenization

Lysate was not

Make sure to incubate the

incubated at room

lysate after the disruption

temperature for

step for 5 minutes at room

5 minutes after the

temperature to allow

disruption step (see

complete dissociation of

step 4 in protocol)

nucleoprotein complexes

RNA is

RNase contamination

Treat all user-made solutions

degraded

of solutions supplied

with DEPC before use (see

by the user

section 3, Maintaining an
RNase-Free Environment)

RNase contamination

See section 3, Maintaining

of plasticware and

an RNase-Free Environment

work station

Frozen tissue samples

Add PureZOL directly to

were allowed to thaw

frozen samples before they

or sit at room

thaw. Do not let starting

temperature

samples sit at room
temperature