7.1.1 Trade Critical Settings

Trade critical settings can affect calibration and/or legal for trade performance. In this document the

symbol

indicates the setting is trade critical. Each time a trade critical setting is altered, the calibration counter will be
incremented by one.

7.2 GSE-LINK

For information on setting up using the GSE-LINK cable refer to GSE-LINK section 5.4.

7.3 Access Full Setup

Full Setup provides access to configure and calibrate the instrument. All items in all menus will be enabled in Full
Setup.

WARNING: Care should be taken to avoid inadvertently altering the Build or Calibration settings.

• Ensure the instrument is on.

• Press and hold the <POWER> and <FUNCTION> keys together for two seconds.

7.4 Access Safe Setup

Safe Setup restricts access to the trade critical settings (marked with

).

• Ensure the instrument is on.

• Press and hold the <POWER> and <ZERO> keys together for two seconds.

7.5 Exit Full or Safe Setup

To save settings, exit setup and return to the normal weighing mode use one of the following methods:

Method 1: Press <POWER> and <FUNCTION> keys together for two seconds.
Method 2: Press <POWER> and <ZERO> together for two seconds.
Method 3: Select - End - from the menus.

Warning:
If the power is interrupted while in setup (i.e. by disconnecting the power cable or pressing the
<POWER> key), unsaved settings will be lost.

7.6 Settings

The following tables identify the settings available for the instrument.