Cooper Instruments & Systems DFI 250X Digital Weight Indicator User Manual

Page 11

Advertising
background image

CF 146

9

V- Rev. 1.0 - 003R-620-100

GROUP

(GRP)

ITEM
(ITM)

SELECT

(SEL)

EDIT
(EDT)

Underline = Defaults

OK

DP

Decimal Point Position

000000, 00000.0, 0000.00, 000.000, 00.0000, 0.00000

Accept

CAP

Maximum Capacity

000100 to 999999 Default = 003000
<SEL>
changes position, <EDT> changes digit.

Accept

RES

Resolution(Count-By)

1, 2, 5, 10, 20, 50, 100

Accept

UNITS

Weighed Units

none, g, kg, lb, t

Accept

HI.RES

Resolution x 10 Mode

OFF, ON

Accept

BUILD

CABLE

4-Wire or 6-Wire

4, 6

Accept

USE

Scale Use

INDUST, OIML, NTEP

Accept

FILTER

Reading Average
(time in seconds)

none, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0

Accept

MOTION

Motion Detection

OFF, 0.5-1.0, 1.0-1.0
Default = 0.5 Graduations per Second

Accept

INIT.Z

Initial Zero on Startup

OFF, ON

Accept

Z.TRAC

Zero Tracking Sensitivity

OFF, SLOW (0.5 grads/sec), FAST (10 grads/sec)

Accept

Z.RANGE

Zero Operating Range

-2_ 2, -1_ 3, -20_20
Default = -2_ 2 (–2% to +2%)

Accept

OPTION

Z.BAND

Zero 'Dead' Band

<SEL> changes position, <EDT> changes digit.
Default = 0

Accept

ZERO

Zero Calibration Routine
Set between zero and full scale
(Current weight displays)

<SEL> key to start. Current weight displays.
Remove all weight.
<SEL>, <EDT> or <OK> starts routine (Z.in P displays).
<ITM> key to exit, <SEL>, <EDT> or <OK> to repeat routine.

SPAN

Span Calibration Routine
Set between zero and full scale
(Current weight displays)

<SEL> key to start. Current weight displays.
Add test weight.
<SEL> or <OK> to show calibration weight value.
Set correct weight. <SEL> changes position, <EDT> changes digit.
<ITM> or <OK> starts routine (S.in P displays).
<ITM> key to exit, <SEL>, <EDT> or <OK> to re-edit calibration weight and repeat
routine.

ED.LIN

Edit Linearization Points
Set between zero and full scale
L1. - - - Select Linearization point
1 to 5
(L1, L2, L3, L4, L5).
(Approx. % of full scale)

<SEL> to step through list of points.
<OK> to change selected point or <ITM> to exit.
After <OK> current weight displays.
Add test weight.
<OK> to enter corrected weight. <SEL> changes position,
<EDT> changes digit.
<OK> starts routine (L.in P displays).
<ITM> key to exit, <OK> to repeat routine.

CAL

CLR.LIN

Clear Linearization Points
L1. - - -
Select Linearization point
1 to 5

<SEL> to step through list of points.
<OK> to select point to clear or <ITM> to exit.
After <OK> press <ITM> to choose Cont. Y (Yes) or Cont. N (No).

Advertising