Cooper Instruments & Systems DFI 250X Digital Weight Indicator User Manual
Page 13

CF 146
11
V- Rev. 1.0 - 003R-620-100
TARG.HI
Target for Overweight
Setpoint 1 (Output 1)
-99999 to 999999
Default = 000000
Accept
TARG.LO
Target for Underweight
Setpoint 2 (Output 2)
-99999 to 999999
Default = 000000
Accept
FORMAT
Date Format Setting
dd.mm.yy or mm.dd.yy
Accept
YEAR
Year Setting
2000 to 2099
<SEL> changes position, <EDT> changes digit.
Accept
MONTH
Month Setting
01 to 12
<SEL> changes position, <EDT> changes digit.
Accept
DAY
Day Setting
01 to 31
<SEL> changes position, <EDT> changes digit.
Accept
HOUR
Hour Setting
00 to 23
<SEL> changes position, <EDT> changes digit.
Accept
CLOC
MINUTE
Minute Setting
00 to 59
<SEL> changes position, <EDT> changes digit.
Accept
SCALE
Scale Base Test Display
Display reads in milliVolts-per-Volt (factory calibrated to 0.1%)
Verify the correct load cell capacity and/or load cell wiring is used. Remove weight
from scale.
TEST
FRC.OUT
Force Outputs
<EDT> advances through outputs (i.e. ON.1 and ON.2)
<OK> turns outputs off and exits test.
Default = OFF
FACTRY
DEFLT
Restore Factory Defaults
Cont. N (No)
Cont. Y (Yes)
Cont. N
Warning: Choosing Cont. Y will clear all stored data except BUILD and CAL
menus.
Accept
- END -
EXIT SETUP
Save settings and return to
normal weighing mode
Accept