1.0 INTRODUCTION

This manual contains information on the installation, calibration and setup of the instrument.

1.1 Approvals (for trade versions only)

NSC approval (4000 divisions at 0.8µV/division).

NMI approval (4000 divisions at 0.8µV/division).

C-tick approved and CE approved.

1.2 Manuals

For more information on this instrument refer to the Reference Manual or Quick Start Manual.

2.0 SHIPPING CONTENTS

The following table identifies the items shipped with the indicator. Please check that your packing box contains the
specified items.

Shipped Items

Other Items (Optional)

• Indicator
• Operator Manual
• Quick Start Manual
• Trade Label (plastic)
• Panel Mount Template
• Function Key Overlay Stickers

• GSE-LINK Cable
• Power Supply
• Desk Mount with Battery Compartment
• U Bracket


3.0 SPECIFICATIONS

Performance
Resolution

Up to 30,000 divisions, minimum of 0.25µV/division, 20 updates/second (Trade
4000 divisions at 0.8µV/division)

Zero Cancellation

±2.0mV/V

Span Adjustment

0.1mV/V to 3.0mV/V full scale

Stability/Drift

Zero: < 0.1µV/°C (+ 8ppm of deadload max)
Span < 8 ppm/°C, Linearity < 20ppm, Noise < 0.2µVp-p

Excitation

5 volts for up to 4 x 350 or 8 x 700 ohm load cells (4-wire or 6-wire plus shield)
Maximum total load cell resistance: 1,000 ohms

A/D Type

24bit Sigma Delta with 8,388,608 internal counts

A/D Conversion Rate

20Hz with FIR filtering > 80dB

Operating Environment

Temperature: 14 to 122°F ambient
Humidity: <90% non-condensing
Storage: –4 to +122°F ambient
IP55 when panel mounted

Case Materials

ABS, Silicon Rubber, Nylon, Acrylic (no halogen used)

Packing Weights

Basic Indicator: 0.75lb

Digital
Display

LED Backlit LCD with six 0.8in high digits with units and annunciators

Setup and Calibration

Full digital with visual prompting in plain messages

Digital Filter

Sliding window average from 0.1 to 4.0 seconds

Zero Range

Adjustable from ±2% to ±20% of full capacity

Power Input
Standard Power Input

12 to 24VDC (2.5 VA max) - ON/OFF key with memory feature

AC

Wall Transformer: 110/240VAC 50/60Hz in 12VDC 0.5A out

Variants

Battery

4 x AA batteries (Alkaline or rechargeable NiMH, NiCad, etc.)

Features
GSE-LINK Data Coupling

Infra-red Connector for optional GSE-LINK PC cable (to RS-232 PC port)

Correction

Five point linearity correction

Outputs

RS-232 automatic transmit, network or printer outputs. Transmission rate: 2400,
4800 or 9600 baud

Assignable Function Key

Unit switching, counting, manual hold, peak hold, live weight and totalizing