6 setting ramp times, 7 setting soak temperature – Humboldt H-3185B Rapid, Freeze Thaw Cabinet User Manual

Page 12

Advertising
background image

12

5.6 Setting Ramp Times

1) Continuing from section 5.5, while the green display is alternating between

“PFrE” and the set option as in the previous section, continue pressing the

Mode button until the green display reads “rt1” (Figure 15). The display

will alternate between “rt1” and the ramp time value (in minutes) for the first

ramp in the cycle. It is capable of 8 ramp operations in the cycle. This is the

first ramp time setting. The soak time and temperature for the first ramp

must be programmed or skipped before going to the next ramp time

setting. It is preset to 5 minutes.

2) To change the ramp time, press the up or down arrow button until the

desired ramp time is achieved.

Note: Due to the thermal behavior of the Freeze/Thaw machine and the

response of the sample material, it is not recommended that this setting be

increased as this could cause fluctuations in the cycle ramp and greatly increase

the overall cycle time.

5.7 Setting Soak Temperatures

1) Continuing from section 5.6, while the green display is alternating between

“rt1” and the set ramp time value, continue pressing the Mode button until

the green display flashes “SL1” (Figure 16). The display will alternate

between “SL1” and the soak temperature value in degrees F. As in the case

of the ramp time, this is the first of 8 soak levels available. Soak level 1 is

preset to 0.

2) To change the soak temperature, press the up or down arrow button until

the desired temperature is achieved.