5.8 Setting Soak Time

1) Continuing from section 5.7, while the green display is alternating between

“SL1” and the soak temperature, continue pressing the Mode button until

the green display flashes “St1” (Figure 17). The display will alternate

between “St1” and the soak time (in minutes). As in the case of the ramp

time, this is the first of 8 soak times available. Soak time 1 is preset to 0 so

that as soon as the temperature reaches the set point, the controller

switches to the next ramp to reach the next temperature set point.

2) To change the soak time, press the up or down arrow button until the

desired time is achieved.

Note: It is necessary to set ramp time 1, soak level 1, and soak time 1 before

setting ramp 2 and soak 2. Simply continue pressing the Mode button until

rt2, SL2, and St2 appear in the green display. If no more temperature levels

are required in the cycle, return to the main operating mode by pressing

and holding down the Menu button until the red display reads the current

temperature. Ramp time 2 is preset to 5 minutes. Soak level 2 is preset to 40°F.

Soak time 2 is preset to 0. The remaining 6 ramp times are preset to 0. If the

ramp time is set to 0, the ramp will be skipped.

6.0 Service

For any service issues, contact Humboldt Scientific, Inc. at 1-800-537-4183

between 8:00 AM and 5:00 PM Eastern Time.