4) Plug the main power cord from the controller box into an appropriate

receptacle. The controller requires a 125VAC, 30 amp power source.

5) Connect the thermocouple wire from the machine to the matching

connector coming from under the chart recorder.

6) The main circuit breaker is located on the side of the controller box near the

door latches. Set the breaker to the ON position.

7) Wait for the controller to go through its self test. When the self test

is complete, the upper red display and the lower green display will show

temperature readings. The green display is the set value and the red display

is the actual temperature reading from the TC (Figure 4).