9. If this is a variable speed version, turn the speed control to MAX.
10. Turn the centrifuge on by pushing the ‘START’ button.
11. The ‘RUN’ indicator light will illuminate.
12. The unit will accelerate to its set speed.
13. Listen to the sound of the centrifuge. A smooth whirring sound should be

heard. If there are any loud or unusual sounds, stop the centrifuge by

pushing the ’OPEN / EMERGENCY STOP’ button immediately and refer

to page 12 on troubleshooting.

14. While the machine is running, try to turn the latch counter-clockwise.

Power may be cut to the motor but you should not be able to fully turn

the latch. If it is possible to turn the latch and open the lid while the unit

is running, contact your authorized dealer or The Drucker Company.

Close and latch the lid.

15. Push the ‘OPEN / EMERGENCY STOP’ button. The ‘RUN’ indicator

light should go out and the motor should slow to a stop.

16. The lid should remain locked until the rotor has stopped. If the machine

unlocks prematurely, contact your authorized dealer or The Drucker

Company. Once the rotor has stopped, a beeper will sound and the

interlock system will become disengaged for sixty (60) seconds. The

‘UNLOCKED’ indicator light will be illuminated during this time.

17. To gain entry into the centrifuge after this period has ended, simply press

the ‘OPEN / EMERGENCY STOP’ button. The lid will unlock for fifteen

(15) additional seconds. Repeat if necessary.

(Continued)

After the centrifuge has passed this procedure,

it is ready for operation.

If you would like to make adjustments to your machine’s

settings, please continue on to “Additional Set-Up Procedures”.

Additional Set-Up Procedures (Optional):

1. Verify the Run-Time Preset:

NOTE: The centrifuge must be plugged in and the lid must be opened to
access programming functions!

a. Push and hold the ‘START’ button for approximately three (3) seconds.

The ’LATCH’ indicator light will begin to flash; indicating program mode.

b. When you release the ‘START’ button, the ‘RUN’ indicator light will begin

to flash. Each flash of the ‘RUN’ indicator light represents one minute of

run time. If the time indicated is not desired, follow the instructions on

the rear of the centrifuge to change it.