BALANCED LOADS

Your centrifuge must contain a balanced load in order to work properly.

To ensure that the load is balanced, keep these rules in mind when

inserting test tube samples:

1. Opposing tube carriers must be identical.
2. Opposing tube carriers must be empty or loaded with an equal number of

equally weighted samples.

3. If an odd number of samples is to be spun, use a water-filled tube to mate

with the unpaired one.

Note: The 755 is designed to detect serious out-of-balance situations and will

safely terminate a run if one is detected. The 755 will cut power to the motor

and wait for the rotor to stop spinning before unlocking. The load should be

balanced before an attempt is made to re-run.

4. When loading tubes into opposing
carriers, the carriers will be balanced
when the tubes are loaded diagonally
from one another. See the illustration
to the right.

Note: When loading tubes that are

longer than 100mm in length, the

maximum capacity is two (2) tubes

per carrier. Place these longer tubes

in the right-hand side of each carrier.

Note: When spinning glass tubes in

the 4-place carrier, it is recommended

that the 0.25” tube cushion (p/n 9150)

be placed in the bottom of the carrier.

755-24 Shown