Ofs.1 – Thermcraft XST-3-0-36-3V User Manual

Page 250

Advertising
background image

User Calibration

Installation and Operation Handbook

7-4

2416 Controller

Set Offset 1

Use

or

to set the offset value of Process Value 1

(PV1).
The offset value is in display units.

Press


The table below shows the parameters which appear after

OFS.1’. These are all read only values and are for information.

Press

to step through them.

mV.1

IP1 measured value (at terminals)

CJC.1

IP1 Cold Junction Compensation

Li.1

IP1 Linearised Value

PV.SL

Not available in Model 2416

If you do not want to look at these parameters, then press

and this returns you to the ‘

iP-LiSt’ header.

To protect the calibration against unauthorised adjustment, return
to Operator level and make sure that the calibration parameters
are hidden. Parameters are hidden using the ‘Edit’ facility
described in Chapter 3, Access Levels.






OFS.1

0

See table on the

right for additional

parameters