Model 7620U Bench Scale User’s Manual

16

Step-by-Step

Instructions for

CONFIG Mode

Follow these steps to access and configure the
items in the CONFIG menu. Refer to Figure 2.

1. From the DIAG display press the ZERO key,

or from normal weighing mode, move Switch
1 to Menu Mode or OPEN position, then
press the ZERO key…

“CONFIG” is displayed.

2. Press the F1 key…

“FILTER” is displayed.

3. Press the F1 key…

The current setting is displayed. Use the
ZERO key to toggle between “FLTR
ON”
and “FLTR OFF”

4. Press the F1 key…

Filter selection is stored.

5. Press the ZERO key…

“BAUD” is displayed.

6. Press the F1 key…

The current setting is displayed. Use the
ZERO key to toggle between the five
choices: “1200, 2400, 4800, 9600, or
19200” baud

7. Press the F1 key…

Baud rate selection is stored.

8. Press the ZERO key…

“PARITY” is displayed.

9. Press the F1 key…

The current setting is displayed. Use the
ZERO key to toggle between the three
choices: “EVEN, ODD, NONE.”

Press the ZERO key to
scroll through lists of
selections.

Press the F1 key to
make a selection

Tip: Quickly and easily
view current scale
configuration directly
from the front panel
without opening the
scale or setting
switches as follows:

During the display
segment test on
power-up, press the
F1 key. The display
will prompt “ABORT”
followed by “BAUD.”
Press the
ZERO key to
scroll through the
choices, or press the
F1 key to view a
current scale configu-
ration.

When you are done,
press the
ZERO key
until “DONE” is
displayed. Press the
F1 key to return to the
normal weighing
mode.