Method RS101 Calibration

Mount the Model 7605/Model 7606 assembly on a non-metallic tripod. This helps
prevent interaction between the assembly and the surroundings, and allows the
operator to be hands-free near the instrumentation.

Block Diagram – Calibration Test Setup

During calibration, keep the assembly several diameters away from metal
objects. Space the current probe and other instrumentation at least one meter
from the assembly to keep it from the influence of the Model 7605.

R

EQUIRED INSTRUMENTATION

Model 7605/Model 7606 assembly

Signal source—see page 26 for more information

Two measuring instruments—see page 27 for more information

Current probe—see page 27 for more information

MIL-STD-462D Method RS101 Testing