Equation 9 – ETS-Lindgren 7606 Coil Antenna/Shielded Coil User Manual

Page 33

Advertising
background image

Stepped scans must dwell at each frequency for one second. f

0

is the

tuned frequency, not the starting frequency. Usually, for analog scans,
the frequency range is broken into octave or decade bands and the
scan rate is changed at each band. The following table shows
suggested bands and analog scan rates. For stepped scans, the
frequencies may be computed using equation 9.

Frequency Band

Analog Scan Rate

Total Scan Time

30 Hz–100 Hz

2 Hz/s

35 s

100 Hz–200 Hz

8 Hz/s

12.5 s

200 Hz–400 Hz

17 Hz/s

11.8 s

400 Hz–700 Hz

32 Hz/s

9.3 s

700 Hz–1.5 kHz

61 Hz/s

13.1 s

1.5 kHz–3 kHz

127 Hz/s

11.8 s

3 kHz–6 kHz

250 Hz/s

12 s

6 kHz–10 kHz

460 Hz/s

8.7 s

10 kHz–20 kHz

850 Hz/s

11.8 s

20 kHz–40 kHz

1.7 kHz/s

11.8 s

40 kHz–100 kHz

3.8 kHz/s

15.9 s

Suggested Scan Rates for Analog Scanning

E

QUATION

9

4.

If susceptibility is noted, select three or more test frequencies per
octave at frequencies where susceptibilities are present. For example,
if five frequencies within an octave show susceptible responses, select
at least three of them for further testing. The three selected should be
those with the maximum indication of susceptibility.

MIL-STD-462D Method RS101 Testing

| 33