Laser Safety

www.ets-lindgren.com

Exposure to any laser radiation during normal operation is not likely because the

EMCenter uses only fiber-coupled lasers. However, for the safety of all

employees, please comply with the following rules:

When the EMCenter is not in use, power off the unit.

The remote interlock connection should be connected to an emergency

master disconnect and in series with the room door or fixture

interlocks.

Connect all fiber cables and install protective covers before switching

on the system. Protective covers are identified by the following label:

Do not attempt to turn on the EMCenter if the fiber optic cables show

any sign of damage.

Safety Features

To make the EMCenter as safe as possible, the primary safety system of each

plug-in card is designed to work standalone, without any intervention from the

EMCenter. Apart from this, the EMCenter has its own safety interlock system,

which is connected in series with the interlock system of every installed card.

Additionally, the following safety measures are implemented:

The EMCenter is equipped with a remote interlock system. For

example, this interlock system would prevent inadvertent laser

radiation by preventing someone from entering a shielded room and

stepping on a fiber.

To prevent accidental activation of the laser, an Acknowledge button

will appear directly after the Laser Start button is pressed. The laser

will start only if this button is pressed within a four-second timeframe.

A visual Laser On indicator warns the operator of laser activity. The

Laser On LED on the front panel of EMCenter indicates that at least

one laser is active.