Pre-Installation Tasks

15

4.0 Pre-Installation Tasks

Before connecting any components, follow the

safety information in the ETS-Lindgren

Product Information Bulletin included with your

shipment.

Bench Test

Perform the following procedures to verify system operation.

R

EADOUT AND

P

ROBE

B

ENCH

T

EST

1.

Remove the plastic caps from the readout's fiber optic connectors.

Remove the protective covers from the fiber optic cable assembly.

Save all protective caps and covers for future use.

2.

Visually inspect the tips of the fiber optic cables to make sure that they

are free of dirt and other contaminants. Connect the fiber optic cable to

the two connectors on the top of the readout; be sure to match the

cable connector colors to those on the readout connectors (yellow to

TRANSMIT; white to RECEIVE).

3.

Connect the other end of the fiber optic cable to the sensor connectors.

Be sure to match the cable connector colors to those on the sensor

connectors (white to XMIT; yellow to RCV).

4.

Set the ARM/OFF switch on the sensor to ARM.